PRODUCT FEATURES | |
| |
4.3'' rapid response capacitive touchscreen | All components and circuits interfaces are isolated from the external environment ro prevent oxidant intrusion |
| |
Long Battery Life | Flow Accuracy |
28.3L/min≥10hours (16.98m⊃3;) 100L/min≥10hours (60m⊃3;) 200L/min≥4hours (48m⊃3;) | ±2.5%(28.3L/min) ±2.5%(100L/min) ±3%(200L/min) |
| |
Compliance with FDA21 CFR Part 11 | |
REMOTE SAMPLING | |
| |
Powerful and stabilized pump to support Remote/Isolator sampling(up to 8m) | |
APPLICATIONS | |
| |
| |
|
|
| |
|
|
SPECIFICATIONS
FLOW RATE: 28.3L/min, 100L/min, 200L/min
FLOW RATE ACCURACY: ±2.5% (28.3L/min), ±2.5%(100L/min), ±3%(200L/min)
ZONE: Over 20000 locations
SAMPLING VOLUMES: 1-6000L, 6 default volumes
DELAY: 0-59 minutes 59 seconds
CYCLES: 1-99cycles
INTERVAL: 0-59 minutes 59 seconds
SAMPLING HEAD: Φ0.60mm*280holes (28.3L/min), Φ0.60mm*280holes (100L/min), Φ0.70mm*400holes (200L/min)
DIMENSIONS: Φ110mm, H: 202mm (without foot mat)
WEIGHT: 2.7kg (6.0lbs)
EXTERNAL SURFACE: Anodized aluminium
POWER: 24V±10% AC adapter
BATTERY: 10.8V, 6800mAh rechargeable lithium battery, running for over 10 hours
DATA STORAGE: 20000 sets of sample records
DISPLAY: 4.3" capacitive touch screen
LANGUAGE: Chinese, English
CALIBRATION FREQUENCY: Recommend once a year
OPERATING CONDITIONS TEMPERATURE:5℃-30℃, relative humidity up to 80%
STORAGE CONDITIONS TEMPERATURE: 0℃-40℃, relative humidity up to 80%
SAFETY: EN 61010-1:2010, EN 61326-1:2013, EN 61326-2-2:2013,EN 61000-6-1:2007, EN 61000-6-3:2007+A1
WARRANTY: 2 years after activation
VIDEO