Waterproof Airborne Particle Counter for Compressed Air

État de disponibilité:

Waterproof Airborne Particle Counter for Compressed Air

Airborne Particle Counter (APC) is fully compliant with ISO14644-2015 and ISO21501-4 standard.

Capacitive touch screen8.0″

Zone20,000

Data 5 million sets

Portable 6.1Kg

Product details

Product Feature

8 inch capacitive touchscreen, quick response

Even with two pairs of latex gloves

Sealed 316L stainless steel enclosure and dust plug to prohibit dust and water

Automatically Collect Air Samples

APC data exceed the limit,triggers the BAS to start sampling

Operation Interface

Data

-Standard report: one key to generate ISO/EUGMPI CHINESEGMP reports

-Data review and data export

1. Sum up data by zone.

2.Freely switch counting unit between △/Σ/#,m /ft.

3.Automatically generate UCL values and average values

4.Search for sample records precisely by time,zone and operator.

-Data integrity: featured with audit trail.

Sampling

- Real-time concentration curves, to get a better understanding of environmental particle condition.

- Quick editing function, to make sampling more flexible.

Setup

- Clear layout for orders. Simple and easy to use.

- One-click for system software upgrade.

- Multiple sets of personalized sampling schemes

can be customized for any zone.


Sampling SCM

Sampling SCM is used for setting up the scheme for sampling. The scheme consists of volume, sample time, delay, etc. It can be added, amended or removed in this menu. The sampling SCM can be associated with zone here. Touch【Sampling】icon of the left menu in the setting interface to set the scheme.


Description

Name

The name of sampling SCM (up to 19 characters). Add: touch the input area to enter a new scheme, touch【】icon to save the settings. Touch【】icon to select an existing scheme from the pull-down menu. Touch【】icon to find a scheme, and touch【OK】icon to confirm. Bulk sampling scheme addition: select the existing sampling scheme from the pull-down menu, modify the name of the scheme and touch【】icon to save, can select between【Cover the original sampling SCM】icon or【Save as a new sampling SCM】icon. The scheme “default” is a system default sampling scheme. It can be modified, but it cannot be deleted. After modifying the name of the Default scheme, touch【】icon to save it as a new scheme.



SPECIFICATIONS



FLOW RATE: A110: 100LPM±3%

A120: 50LPM±3%


A130: 28.3LPM±3%

SIZE RANGE: A110: 0.5µm-25µm

A120, A130: 0.3µm-25µm


SIZE CHANNELS: A110: 0.5µm, 1.0µm, 2.0µm, 3.0µm, 5.0µm, 10.0µm

A120, A130: 0.3µm, 0.5µm, 1.0µm, 3.0µm, 5.0µm, 10.0µm

SIZE RESOLUTION: <15% @ 0.5&micro;m (meets ISO 21501-4)

COUNT EFFICIENCY: A110: 50% &plusmn;20% for 0.5&micro;m, 100% &plusmn;10% for >0.75&micro;m (meets ISO 21501-4 and JIS B9921)

A120, A130: 50% &plusmn;20% for 0.3&micro;m, 100% &plusmn;10%for >0.45&micro;m (meets ISO 21501-4 and JIS B9921)

CONCENTRATION LIMIT: A110: 560,000 particles/ft&sup3;at 10% coincidence loss

A120: 840,000 particles/ft&sup3;at 10% coincidence loss

A130: 1,400,000 particles/ft&sup3;at 10% coincidence loss

ZERO COUNT: <1count/5min

EXHAUST: Internal HEPA filter (>99.997%@0.3&micro;m)

ZONE/ LOCATION: 20,000 rooms and 200,000 locations

SAMPLING TIME: 1 second to 168 hours

CYCLES: 1000 samples on one location

LASER SOURCE: Long life laser

DATA STORAGE: 3.5G, ≥5,000,000 sets of sample records


1

1

sur: 
En vertu d'un: 

modèle du produit

le contenu est vide!

NOUS CONTACTER
Retour d'information
MicronView Limited tire parti de l'optique et de la biotechnologie pour atteindre une surveillance de l'environnement stérile rapide et en temps réel, en fournissant des instruments de surveillance avancés et en consommables pour les environnements de production et les produits finis des industries alimentaires, pharmaceutiques et cosmétiques.
Navigation Rapide

Catégorie de produits

MicronView Limited

1-520-528-1882
mv@micronview.com
2300 Highway 365 Suite 110, Pays-Bas Texas 77630, États-Unis
droits d'auteur 2021 Micron View Limited. Tous les droits sont réservés. Plancher| Soutenu par En avance