Autoclavable Airborne Particle Counter

Airborne Particle Counter (APC) is fully compliant with ISO14644-2015 and ISO21501-4 standard.

Capacitive touch screen8.0″

Zone20,000

Data 5 million sets

Portable 6.1Kg

Product details

Product Feature

8 inch capacitive touchscreen, quick response

Even with two pairs of latex gloves

Sealed 316L stainless steel enclosure and dust plug to prohibit dust and water

Automatically Collect Air Samples

APC data exceed the limit,triggers the BAS to start sampling

Operation Interface

Data

-Standard report: one key to generate ISO/EUGMPI CHINESEGMP reports

-Data review and data export

1. Sum up data by zone.

2.Freely switch counting unit between △/Σ/#,m /ft.

3.Automatically generate UCL values and average values

4.Search for sample records precisely by time,zone and operator.

-Data integrity: featured with audit trail.

Sampling

- Real-time concentration curves, to get a better understanding of environmental particle condition.

- Quick editing function, to make sampling more flexible.

Setup

- Clear layout for orders. Simple and easy to use.

- One-click for system software upgrade.

- Multiple sets of personalized sampling schemes

can be customized for any zone.

Device

Device is used for displaying the information of the software and firmware. It includes device name, model, serial number, hardware version, software version, etc. The software can be upgraded on this interface. See chart 5-3-13.

Update is used for upgrading the software version. APC supports continuous version upgrades and cross version upgrades. Insert the MicronView USB flash drive into the USB A port at the back of the device. Touch the【Update】icon, search and select the upgrade file, touch【】icon to confirm. See chart 5-3-14.

Back- up is used to export all information in the device to the MicronView USB flash drive in the form of db files. When connected to the USB flash drive, touch【Back-up】icon, and a prompt box ‘Backup completed’ will display after the back-up is complete. Touch【OK】icon to confirm. Restore is used to import device information in the form of db file to the device. Insert the MicronView USB flash drive with the db file includes device information into the USB A port at the back of the device. Touch【Restore】icon to restore the device information. When restoring, it is necessary to confirm that the serial number and version of the db file name in the USB flash drive are the same as those displayed in the device information; otherwise, the prompt message of ‘restoration failure’ will be displayed. After the restoration is completed, the device automatically shut down. After restarting, the device can be used normally.



SPECIFICATIONS



FLOW RATE: A110: 100LPM±3%

A120: 50LPM±3%


A130: 28.3LPM±3%

SIZE RANGE: A110: 0.5µm-25µm

A120, A130: 0.3µm-25µm


SIZE CHANNELS: A110: 0.5µm, 1.0µm, 2.0µm, 3.0µm, 5.0µm, 10.0µm

A120, A130: 0.3µm, 0.5µm, 1.0µm, 3.0µm, 5.0µm, 10.0µm

SIZE RESOLUTION: <15% @ 0.5&micro;m (meets ISO 21501-4)

COUNT EFFICIENCY: A110: 50% &plusmn;20% for 0.5&micro;m, 100% &plusmn;10% for >0.75&micro;m (meets ISO 21501-4 and JIS B9921)

A120, A130: 50% &plusmn;20% for 0.3&micro;m, 100% &plusmn;10%for >0.45&micro;m (meets ISO 21501-4 and JIS B9921)

CONCENTRATION LIMIT: A110: 560,000 particles/ft&sup3;at 10% coincidence loss

A120: 840,000 particles/ft&sup3;at 10% coincidence loss

A130: 1,400,000 particles/ft&sup3;at 10% coincidence loss

ZERO COUNT: <1count/5min

EXHAUST: Internal HEPA filter (>99.997%@0.3&micro;m)

ZONE/ LOCATION: 20,000 rooms and 200,000 locations

SAMPLING TIME: 1 second to 168 hours

CYCLES: 1000 samples on one location

LASER SOURCE: Long life laser

DATA STORAGE: 3.5G, ≥5,000,000 sets of sample records

DISPLAY: 8” LCD capacitive touch screen

LANGUAGE: Chinese, English

DATA RELIABILITY: Compliant with 21 CFR Part 11

REPORTS: Compliant with ISO/EUGMP/CHINESEGMP/Fed Std

CALIBRATION FREQUENCY: Recommend at least once a year

DIMENSIONS (HXWXD): 254x204x263mm

WEIGHT: 6.1Kg (13.45lbs) (without battery)

ENCLOSURE: 316L Stainless Steel

POWER: AC 100-240V, 50 Hz/60 Hz

BATTERY: 10.8V, 9000mAh, rechargeable lithium battery

BATTERY ENDURANCE: A110: running for 4 hours

A120: running for 7 hours


A130: running for 9 hours

MAX POWER: consumption AC 90W

OPERATING CONDITIONS TEMPERATURE:: 5℃-35℃, relative humidity up to 95%

STORAGE CONDITIONS TEMPERATURE: -10℃-40℃, relative humidity up to 98%

SAFETY: EN 61010-1:2010, EN 61326-1:2013, EN 61326-2-2:2013,EN 60825-1:2014, EN 61000-6-1:2007, EN 61000-6-3:2007+A1,FCC Part 15, Subpart B

WARRANTY: 2 years after activation






Anterior: 
Siguiente: 
CONTÁCTENOS
Realimentación
MicronView Limited se aprovecha de la óptica y la biotecnología para lograr un monitoreo de entorno estéril rápido y en tiempo real, proporcionando instrumentos de monitoreo avanzados y consumibles para entornos de producción y productos terminados en las industrias alimentaria, farmacéutica y cosmética.
Navegación rápida

Categoría de Productos

MicronView Limited

1-520-528-1882
mv@micronview.com
2300 Highway 365 Suite 110, Nederland Texas 77630,  USA.
Derechos de autor 2021 Micron View Limited. Todos los derechos reservados. Mapa del sitio| Con apoyo de Letrero