Vous êtes ici: Maison » Des produits » Digital Airborne Particle Counter for Cleanroom

Les détails du produit

loading

Partager sur:
facebook sharing button
twitter sharing button
line sharing button
wechat sharing button
linkedin sharing button
pinterest sharing button
sharethis sharing button

Digital Airborne Particle Counter for Cleanroom

État de disponibilité:

Digital Airborne Particle Counter for Cleanroom


Airborne Particle Counter (APC) is fully compliant with ISO14644-2015 and ISO21501-4 standard.

Capacitive touch screen8.0″

Zone20,000

Data 5 million sets

Portable 6.1Kg

Product details

Product Feature

8 inch capacitive touchscreen, quick response

Even with two pairs of latex gloves

Sealed 316L stainless steel enclosure and dust plug to prohibit dust and water

Automatically Collect Air Samples

APC data exceed the limit,triggers the BAS to start sampling

Operation Interface

Data

-Standard report: one key to generate ISO/EUGMPI CHINESEGMP reports

-Data review and data export

1. Sum up data by zone.

2.Freely switch counting unit between △/Σ/#,m /ft.

3.Automatically generate UCL values and average values

4.Search for sample records precisely by time,zone and operator.

-Data integrity: featured with audit trail.

Sampling

- Real-time concentration curves, to get a better understanding of environmental particle condition.

- Quick editing function, to make sampling more flexible.

Setup

- Clear layout for orders. Simple and easy to use.

- One-click for system software upgrade.

- Multiple sets of personalized sampling schemes

can be customized for any zone.



SPECIFICATIONS



FLOW RATE: A110: 100LPM±3%

A120: 50LPM±3%


A130: 28.3LPM±3%

SIZE RANGE: A110: 0.5µm-25µm

A120, A130: 0.3µm-25µm


SIZE CHANNELS: A110: 0.5µm, 1.0µm, 2.0µm, 3.0µm, 5.0µm, 10.0µm

A120, A130: 0.3µm, 0.5µm, 1.0µm, 3.0µm, 5.0µm, 10.0µm

SIZE RESOLUTION: <15% @ 0.5&micro;m (meets ISO 21501-4)

COUNT EFFICIENCY: A110: 50% &plusmn;20% for 0.5&micro;m, 100% &plusmn;10% for >0.75&micro;m (meets ISO 21501-4 and JIS B9921)

A120, A130: 50% &plusmn;20% for 0.3&micro;m, 100% &plusmn;10%for >0.45&micro;m (meets ISO 21501-4 and JIS B9921)

CONCENTRATION LIMIT: A110: 560,000 particles/ft&sup3;at 10% coincidence loss

A120: 840,000 particles/ft&sup3;at 10% coincidence loss

A130: 1,400,000 particles/ft&sup3;at 10% coincidence loss

ZERO COUNT: <1count/5min

EXHAUST: Internal HEPA filter (>99.997%@0.3&micro;m)

ZONE/ LOCATION: 20,000 rooms and 200,000 locations

SAMPLING TIME: 1 second to 168 hours



Description

Area    

The area of the sampling room. The area can choose not to be entered. If the area is not entered, it will be displayed as N/A. Standard reports cannot be generated without entering area.

Standards    

Select the standard from the pull-down menu. The standards include ISO 14644-1: 1999, ISO 14644-1: 2015, EUGMP-ISO: 1999, EUGMP-ISO: 2015, Chinese GMP 2010-ISO: 1999, Chinese GMP 2010-ISO: 2015, Fed Std 209E F, Fed Std 209E M. The standard can choose not to be entered, when it is not selected, it will be displayed as N/A. Standard reports cannot be generated without selecting a standard.


1

1

sur: 
En vertu d'un: 
NOUS CONTACTER
Retour d'information
MicronView Limited tire parti de l'optique et de la biotechnologie pour atteindre une surveillance de l'environnement stérile rapide et en temps réel, en fournissant des instruments de surveillance avancés et en consommables pour les environnements de production et les produits finis des industries alimentaires, pharmaceutiques et cosmétiques.
Navigation Rapide

Catégorie de produits

MicronView Limited

1-520-528-1882
mv@micronview.com
2300 Highway 365 Suite 110, Pays-Bas Texas 77630, États-Unis
droits d'auteur 2021 Micron View Limited. Tous les droits sont réservés. Plancher| Soutenu par En avance