Digital Airborne Particle Counter for Cleanroom
Airborne Particle Counter (APC) is fully compliant with ISO14644-2015 and ISO21501-4 standard.
Capacitive touch screen8.0″
Zone20,000
Data 5 million sets
Portable 6.1Kg
Product details
Product Feature
8 inch capacitive touchscreen, quick response
Even with two pairs of latex gloves
Sealed 316L stainless steel enclosure and dust plug to prohibit dust and water
Automatically Collect Air Samples
APC data exceed the limit,triggers the BAS to start sampling
Operation Interface
Data
-Standard report: one key to generate ISO/EUGMPI CHINESEGMP reports
-Data review and data export
1. Sum up data by zone.
2.Freely switch counting unit between △/Σ/#,m /ft.
3.Automatically generate UCL values and average values
4.Search for sample records precisely by time,zone and operator.
-Data integrity: featured with audit trail.
Sampling
- Real-time concentration curves, to get a better understanding of environmental particle condition.
- Quick editing function, to make sampling more flexible.
Setup
- Clear layout for orders. Simple and easy to use.
- One-click for system software upgrade.
- Multiple sets of personalized sampling schemes
can be customized for any zone.
SPECIFICATIONS
FLOW RATE: A110: 100LPM±3%
A120: 50LPM±3%
A130: 28.3LPM±3%
SIZE RANGE: A110: 0.5µm-25µm
A120, A130: 0.3µm-25µm
SIZE CHANNELS: A110: 0.5µm, 1.0µm, 2.0µm, 3.0µm, 5.0µm, 10.0µm
A120, A130: 0.3µm, 0.5µm, 1.0µm, 3.0µm, 5.0µm, 10.0µm
SIZE RESOLUTION: <15% @ 0.5µm (meets ISO 21501-4)
COUNT EFFICIENCY: A110: 50% ±20% for 0.5µm, 100% ±10% for >0.75µm (meets ISO 21501-4 and JIS B9921)
A120, A130: 50% ±20% for 0.3µm, 100% ±10%for >0.45µm (meets ISO 21501-4 and JIS B9921)
CONCENTRATION LIMIT: A110: 560,000 particles/ft³at 10% coincidence loss
A120: 840,000 particles/ft³at 10% coincidence loss
A130: 1,400,000 particles/ft³at 10% coincidence loss
ZERO COUNT: <1count/5min
EXHAUST: Internal HEPA filter (>99.997%@0.3µm)
ZONE/ LOCATION: 20,000 rooms and 200,000 locations
SAMPLING TIME: 1 second to 168 hours
Description
Area
The area of the sampling room. The area can choose not to be entered. If the area is not entered, it will be displayed as N/A. Standard reports cannot be generated without entering area.
Standards
Select the standard from the pull-down menu. The standards include ISO 14644-1: 1999, ISO 14644-1: 2015, EUGMP-ISO: 1999, EUGMP-ISO: 2015, Chinese GMP 2010-ISO: 1999, Chinese GMP 2010-ISO: 2015, Fed Std 209E F, Fed Std 209E M. The standard can choose not to be entered, when it is not selected, it will be displayed as N/A. Standard reports cannot be generated without selecting a standard.