LCD Airborne Particle Counter for Lab
Time
Time is used for setting up the system time on the device. Touch the input box on the right of the date, enter the current date. Touch the pull-down menu, select the date format of YY-MM-DD, DD-MM-YY, MM-DD-YY. Touch【】icon to save the settings. The setting of the time is the same as the date.
SPECIFICATIONS
FLOW RATE: A110: 100LPM±3%
A120: 50LPM±3%
A130: 28.3LPM±3%
SIZE RANGE: A110: 0.5µm-25µm
A120, A130: 0.3µm-25µm
SIZE CHANNELS: A110: 0.5µm, 1.0µm, 2.0µm, 3.0µm, 5.0µm, 10.0µm
A120, A130: 0.3µm, 0.5µm, 1.0µm, 3.0µm, 5.0µm, 10.0µm
SIZE RESOLUTION: <15% @ 0.5µm (meets ISO 21501-4)
COUNT EFFICIENCY: A110: 50% ±20% for 0.5µm, 100% ±10% for >0.75µm (meets ISO 21501-4 and JIS B9921)
A120, A130: 50% ±20% for 0.3µm, 100% ±10%for >0.45µm (meets ISO 21501-4 and JIS B9921)
CONCENTRATION LIMIT: A110: 560,000 particles/ft³at 10% coincidence loss
A120: 840,000 particles/ft³at 10% coincidence loss
A130: 1,400,000 particles/ft³at 10% coincidence loss
ZERO COUNT: <1count/5min
EXHAUST: Internal HEPA filter (>99.997%@0.3µm)
ZONE/ LOCATION: 20,000 rooms and 200,000 locations
SAMPLING TIME: 1 second to 168 hours
CYCLES: 1000 samples on one location
LASER SOURCE: Long life laser
DATA STORAGE: 3.5G, ≥5,000,000 sets of sample records
DISPLAY: 8” LCD capacitive touch screen
LANGUAGE: Chinese, English
DATA RELIABILITY: Compliant with 21 CFR Part 11
REPORTS: Compliant with ISO/EUGMP/CHINESEGMP/Fed Std
CALIBRATION FREQUENCY: Recommend at least once a year
DIMENSIONS (HXWXD): 254x204x263mm
WEIGHT: 6.1Kg (13.45lbs) (without battery)
ENCLOSURE: 316L Stainless Steel
POWER: AC 100-240V, 50 Hz/60 Hz
BATTERY: 10.8V, 9000mAh, rechargeable lithium battery
BATTERY ENDURANCE: A110: running for 4 hours
A120: running for 7 hours
A130: running for 9 hours
MAX POWER: consumption AC 90W
OPERATING CONDITIONS TEMPERATURE:: 5℃-35℃, relative humidity up to 95%
STORAGE CONDITIONS TEMPERATURE: -10℃-40℃, relative humidity up to 98%
SAFETY: EN 61010-1:2010, EN 61326-1:2013, EN 61326-2-2:2013,EN 60825-1:2014, EN 61000-6-1:2007, EN 61000-6-3:2007+A1,FCC Part 15, Subpart B
WARRANTY: 2 years after activation
1