You are here: Home » Products » Hide Products » 316 Stainless Steel Particle Counter

Product Detail

Share to:

316 Stainless Steel Particle Counter

Availability:

316 stainless Steel Particle Counter

Airborne Particle Counter (APC) is fully compliant with ISO14644-2015 and ISO21501-4 standard.

Capacitive touch screen8.0″

Zone20,000

Data 5 million sets

Portable 6.1Kg

Product details

Product Feature

8 inch capacitive touchscreen, quick response

Even with two pairs of latex gloves

Sealed 316L stainless steel enclosure and dust plug to prohibit dust and water

Automatically Collect Air Samples

APC data exceed the limit,triggers the BAS to start sampling



Troubleshooting

5



Battery cannot be charged when AC adapter is working properly.



Poor battery contact.

Check and plug in the battery again.

Battery failure.

Contact the factory to replace the battery.

Charging port failure.

Return to us or the local distributor for service.

6


Battery cannot be fully charged when AC adapter supply.


Battery aging.

Contact the factory to replace the battery.

Internal parts failure.

Return to us or the local distributor for service.

7

Sudden power failure when AC adapter is connected.

AC adapter failure.

If the problem persists after replacing another AC adapter, return to us for service.

8


Device suddenly power off when battery is connected.


Low battery.

Plug in the AC adapter and restart the device.

Battery failure.

Contact the factory to replace the battery.




SPECIFICATIONS



FLOW RATE: A110: 100LPM±3%

A120: 50LPM±3%


A130: 28.3LPM±3%

SIZE RANGE: A110: 0.5µm-25µm

A120, A130: 0.3µm-25µm


SIZE CHANNELS: A110: 0.5µm, 1.0µm, 2.0µm, 3.0µm, 5.0µm, 10.0µm

A120, A130: 0.3µm, 0.5µm, 1.0µm, 3.0µm, 5.0µm, 10.0µm

SIZE RESOLUTION: <15% @ 0.5&micro;m (meets ISO 21501-4)

COUNT EFFICIENCY: A110: 50% &plusmn;20% for 0.5&micro;m, 100% &plusmn;10% for >0.75&micro;m (meets ISO 21501-4 and JIS B9921)

A120, A130: 50% &plusmn;20% for 0.3&micro;m, 100% &plusmn;10%for >0.45&micro;m (meets ISO 21501-4 and JIS B9921)

CONCENTRATION LIMIT: A110: 560,000 particles/ft&sup3;at 10% coincidence loss

A120: 840,000 particles/ft&sup3;at 10% coincidence loss

A130: 1,400,000 particles/ft&sup3;at 10% coincidence loss

ZERO COUNT: <1count/5min

EXHAUST: Internal HEPA filter (>99.997%@0.3&micro;m)

ZONE/ LOCATION: 20,000 rooms and 200,000 locations

SAMPLING TIME: 1 second to 168 hours

CYCLES: 1000 samples on one location

LASER SOURCE: Long life laser

DATA STORAGE: 3.5G, ≥5,000,000 sets of sample records

DISPLAY: 8” LCD capacitive touch screen

LANGUAGE: Chinese, English

DATA RELIABILITY: Compliant with 21 CFR Part 11

REPORTS: Compliant with ISO/EUGMP/CHINESEGMP/Fed Std

CALIBRATION FREQUENCY: Recommend at least once a year

DIMENSIONS (HXWXD): 254x204x263mm

WEIGHT: 6.1Kg (13.45lbs) (without battery)

ENCLOSURE: 316L Stainless Steel

POWER: AC 100-240V, 50 Hz/60 Hz

BATTERY: 10.8V, 9000mAh, rechargeable lithium battery

BATTERY ENDURANCE: A110: running for 4 hours

A120: running for 7 hours


A130: running for 9 hours

MAX POWER: consumption AC 90W

OPERATING CONDITIONS TEMPERATURE:: 5℃-35℃, relative humidity up to 95%

STORAGE CONDITIONS TEMPERATURE: -10℃-40℃, relative humidity up to 98%

SAFETY: EN 61010-1:2010, EN 61326-1:2013, EN 61326-2-2:2013,EN 60825-1:2014, EN 61000-6-1:2007, EN 61000-6-3:2007+A1,FCC Part 15, Subpart B

WARRANTY: 2 years after activation






Previous: 
Next: 

PRODUCT CATEGORY

CONTACT US
Feedback
MicronView Limited takes advantage of optics and biotechnology to achieve fast, real-time sterile environment monitoring, by providing advanced monitoring instruments and consumable for production environments and finished products of food, pharmaceutical and cosmetic industries.
Quick Navigation

Products Category

Micronview Limited

  1-520-528-1882
 mv@micronview.com
  2300 Highway 365 Suite 110,  Nederland Texas 77627, USA
Copyright  2021 Micron View Limited.   All rights reserved.  Sitemap | Support By Leadong